4 dosing, Dosing – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 103

Advertising
background image

- 103 -

<Max>: mass higher than the value

of the high threshold

18.4

Dosing

The dosing is the mode which consits in the process of sample weighing

until target mass is reached. The target mass is defined along with dosing

tolerance. Tolerance value is set as a percentage of the target mass,

entering the value percentage.
Example:
Target weight = 100.000g
Tolerance = 2,5% (2,5% of 100g, which amounts to 2,5g)
i.e.: the software accepts correctly dosed value within the following
thresholds: from 97,500g to 102,500g.


F button shortcuts

The function enables the option of quick access for weighing functions,

which are available when buttons F1, F2, F3 or F4 are pressed.

In the mode <DOSING> the following options are available, which a user

can assign to each of the F buttons:

NONE

– no function assigned to the button


ENTER SAMPLE

– function initiates the option of entering the reference

mass along with tolerance (one procedure)

SELECT PRODUCT

– function initiates the option of selecting a product

from product database


LOG IN

– function initiates the option of selecting and logging-in of a

balance’s user

ENTER TARE

– function initiates the option of inserting packaging mass


SELECT TARE

– function initiates the option of selecting packaging

mass from product database


PRINT HEADER

– function initiates the print-out of a designed header


PRINT FOOTER

– function initiates the print-out of a designed footer


VARIABLE 1

– function initiates the selection and editing of variable 1

Advertising