4 additional display, 5 external buttons, Additional display – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 154: External buttons

Advertising
background image

- 154 -

set the balance parameters related to the cooperation with the bar

code reader

<P4.3.1 PORT> - choice of the port, to which the bar code reader is

to be connected

o

accessible options: NONE, COM 1, COM 2

20.4

Additional display

The submenu <ADDITIONAL DISP.> includes the balance settings for

cooperation with the external additional display.

Procedure:

press

button

enter menu <P4 DEVICES>

and then enter in menu group<P4.4 ADDITIONAL DISP.>

set the balance parameters related to the cooperation with the

additional display

<P4.4.1 PORT> - choice of the port, to which the additional display

is to be connected

o

accessible options: none, COM 1, COM 2

CAUTION:
The balance cooperates with an additional display manufactured by
RADWAG.
Ensuring correct cooperation between the balance and the additional
display requires that the parameter of the baud rate value set to 115200
bit/s for the port to which the additional display is plugged.

20.5

External buttons

Submenu <EXTERNAL BUTTONS> contains the settings that enable the

balance to cooperate with external buttons: TARE and PRINT.

The procedure is following:

Press

Enter menu <P4 DEVICES>

Enter in the menu group <P4.5 EXTERNAL BUTTONS>

Activate the buttons

<P4.5.1. TARE> for <YES>