UNPACKING, ASSEMBLING AND START-UP

4.1

Assembling and place of use

The balance should be stored and used in locations free of vibrations and

shakes, free of air movement and dust,

ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C,

ambient relative humidity should not exceed 80%,

during balance operation, ambient temperature in the weighing room should

not change more than 3°C within one hour,

the balance should be located on a stable wall console desk or a stable

working table which is not affected by vibrations and distant from heat

sources,

take special precaution when weighing magnetic objects, as part of the

balance is a strong magnet. Should such loads be weighed, use under hook

weighing option, which removes the weighed load from area influenced by

the balance’s magnet. For assembling the hook for under hook weighing

option see the bottom section of balance’s housing,

in order to avoid the influence of static electricity on the measurement

process, ground the balance’s housing. The grounding bolt is located at the

back of the balance’s housing.

4.2

Unpacking

Cut the sealing tape. Take the balance out of the packaging. Open the box with

accessories and take the items necessary for correct balance operation. Upon

placing the balance in the desired location assembly its weighing pan and other

parts in accordance with the following instruction.