3 commands description, Commands description – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 162

Advertising
background image

- 162 -

22.3

Commands description

Zero the balance

Format: Z CR LF

Accessible responses:

Z_A CR LF

Z_D CR LF

- command understood and in progress

- command carried out

Z_A CR LF

Z_^ CR LF

- command understood and in progress

- command understood but zeroing range exceeded

Z_A CR LF

Z_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement result

Z_I CR LF

- command understood but not accessible at this moment


Tare the balance

Format: T CR LF

Accessible responses:

T_A CR LF

T_D CR LF

- command understood and in progress

- command carried out

T_A CR LF

T_v CR LF

- command understood and in progress

- command understood but taring range exceeded

T_A CR LF

T_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable measurement result

T_I CR LF

- command understood but not accessible at this moment

Give tare value

Format: OT CR LF

Accessible response: OT_TARA CR LF – command carried out

Response format:

1

2

3

4-12

13

14

15

16

17

18

19

O

T

space

tare

space

unit

space

CR

LF

Tare

- 9 characters with right justification

Unit

- 3 characters with left justification

CAUTION:
Tare value is always given in adjustment unit.


Set tare

Format: UT_TARA CR LF, where TARE – tare value

Accessible responses:

UT_OK CR LF - command carried out

UT_I CR LF

- command understood but not accessible at this moment

ES CR LF

- command not recognised (tare format incorrect)

CAUTION:
Use dot in tare format as decimal point.