10.4.2. Value release

Due to various ambient conditions at workplaces, and in order to set balance to

present conditions, it is necessary to determine the value release parameter

as: FAST.+REL., FAST or RELIABLE. Depending on selected option,

weighing time is either shorter or longer.

10.4.3. Autozero function

In order to ensure balance’s precise mass indication, autozero (Auto) software

parameter has been introduced. The application of this function is automatic

control and correction of zero indication. When the function is enabled, it

compares balance indications at declared time interval e.g. 1s, on condition

that weighing pan is unloaded and display indication is close to zero. If results

vary less than declared AUTOZERO range e.g. one division, balance will zero

automatically, marker of stable measurement result–

and precise zero

marker –

will be displayed.

If AUTOZERO function is enabled, then each weighing process starts from

precise zero point. There are, however, some case when this function can be a

disturbing factor in the measuring process; for instance very slow placing of

load on the weighing pan (e.g. load pouring) in such case zero indication

correction can also correct actual indication of loaded mass.