12.11 GLP autotest

AUTOTEST function is designed to aid a user in assessing balance’s operation and

diagnosing the reasons for occurrence of errors in weighing which exceed the

maximum permissible values for a given balance model.

By means of a simple, repeatable and fully documented way the function enables

optimizing balance’s settings to maintain the best possible repeatability and

weighing time at workstation. The main purpose of the function is the possibility of

monitoring the above-mentioned parameters at optional moment and saving

records from the carried out tests in the form of printed reports of the tests that are

automatically generated at the end of examination.

The test controls repeatability of placing the interval weight and determining error of

indication with reference to balance’s maximum capacity.

Testing procedure:

two times loading of the internal weight, followed by 10 times loading of

the internal weight

balance internal adjustment

calculating the value of standard deviation

report printout

Test results:

*Deviation for Max.

*Value of repeatability of indication expressed as standard deviation