Adjustment triggered by elapsing time. It is possible to declare time

intervals which are criteria for balance adjustment. Accessible

settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours

since last carried out adjustment procedure,

CAUTION

Declaring time intervals for balance adjustment is available only in
non-verified balances. In case of verified balances the time interval
between the following adjustments is set for 3 hours.

14.2

Menu adjustment settings

P1 CALIBRATION

P1.1 INT.CALIB.

|

[internal adjustment]

P1.2 EXT.CALIB.

|

[external adjustment]

P1.3 USER CALIB.

|

[user adjustment]

P1.4 CALIB. TEST

|

[adjustment test]

P1.5 AUTO. CAL. T. | NONE/TIME/TEMP./BOTH

[automatic

adjustment]
P1.6 AUTO CAL. C. |

[time of automatic adjustment]

P1.1

INT. CALIB.

– Internal adjustment

Start of internal adjustment process. Adjustment is conducted automatically,

and no user assistance is required. If weighing pan of the balance is loaded,

balance will display a command to unload the weighing pan. If weighing pan is

loaded with a relatively small mass (up to 4% of balance max capacity) then

adjustment process will be carried out automatically, but measuring results

may differ by the amount that was present on balance weighing pan.