External adjustment – RADWAG PS 510.R2.CT Carat Balance User Manual

Page 64

Advertising
background image

- 64 -


CAUTION

- Press Esc button to abort adjustment process.
-If, during adjustment process, weighing pan of a balance is loaded, the

display will show an error message. Adjustment will be stopped. As
weighing pan is unloaded, the adjustment process will continue and be
completed.

14.3.2. External adjustment

The external adjustment in balances NPS series should be carried out with

external mass standard / weigh class F

1

Start of external adjustment process,

Balance will display a command ordering unloading of weighing pan
<REMOVE MASS>

(weighing pan must be empty). When weighing pan is

unloaded, press

button.

The balance determines mass of empty pan and message <CALIBRATION>

is displayed in the bottom line. It is followed by message message <PLACE
MASS

> in the bottom bar, and in the main window the mass value to be

placed on the weighing pan e.g. 200.000g (depending on the type of

balance). Place an external adjustment weight with mass determined on

balance’s display and press

button. The balance determines the mass

and message <CALIBRATION> is displayed in the bottom line. On

completing adjustment process the balance returns to submenu P1.2
INT.CALIB.

If a balance is verified, it is not possible for a user to carry out

external adjustment process.