14.3.3. User adjustment

The external adjustment in balances NPS series should be carried out with

external mass standard / weigh class F

1

Start of external adjustment process,

The first procedure of the process is to enter the mass of the weight that is to

be used for adjustment. The mass must be ≥ 30% Max capacity.

Once the mass of the weight is entered and confirmed, the message is

displayed which asks you to remove the weight from the pan <REMOVE

MASS>

(weighing pan must be empty). After that, press

button, the

balance determines the weight of unloaded pan and message

<CALIBRATION> is shown in the bottom line .

Next, message <PLACE MASS>is displayed in the bottom bar, and in the

main window the value of the mass to be placed e.g. 200.000g (depending

on balance type). Place the weight of the indicated mass and press

button. The balance determines the mass and message <CALIBRATION> is

displayed in the bottom line. After the adjustment process, the balance

returns to submenu P1.2 EXT.CALIB.

If a balance is verified, it is not possible for a user to carry out

external adjustment process.