Dynaflow, User manual - maintenance – Ransburg DynaFlow User Manual User Manual

Page 72

Advertising
background image

LN-9400-00.9

68

DynaFlow

TM

User Manual - Maintenance

forced ON. If the desired flow is achieved while

in this mode of operation, the fluid restriction is

located further downstream to the applicator.

4. Flow Meter Feedback

- Is flow being indicated on the controller?

If it is obvious that material is flowing through

the system (from the GUN or calibration ports if

used), but the controller indicates no flow, then

this indicates that the controller is not detecting

flow meter pulses. This could be caused by one

of the following:

Flow meter gear stuck

Bad flow meter sensor

Disconnected or damaged fiber-optic

cable

Bad fiber-optic receiver

Bad input on the CHANNEL MODULE

Fiber-optic flow meter transmitter battery

has expired. Life expectancy is two years

minimum.

Corrective Action:

Flow Meter: Disassemble, clean, and inspect

the flow meter. Refer to instructions supplied with

the meter.

Try connecting the flow meter in question to a fiber-

optic cable from a properly functioning CHANNEL.

If fluid flow is now observed, then the flow meter

and pickup sensor are functioning properly.

Fiber-Optic Cable: Simply shine a light into one

end of the cable and have someone observe the

opposite end. If light is visible through the cable

then it is probably good, although it is possible

that the cable is damaged and will not conduct

enough light to operate properly.

Fiber-Optic Detector: With fluid flowing, observe

the LED located on the fiber-optic receiver. The

LED is an indication that fluid is flowing and pulses

are being received from the flow meter. If the LED

is not ON, then replace or swap the fiber-optic

receiver and try again.

If everything checks good but no fluid flow is dis-

played, replace the CHANNEL MODULE. Make

sure that the hardware settings located on the

module are configured identically to the one be-

ing replaced.

5. Transducer Failure

Corrective Action:

Perform the following with the GUN OFF and

calibration valves closed. Place a pressure gage

(if one does not already exist) in the pilot line be-

tween the transducer and the fluid regulator. Make

sure that the air supply pressure to the transducer

is at least 100 psi. Go to the FORCE I/O menu for the

CHANNEL in question. Force Control Output volt-

age (or 4-20 ma) to several different values and

observe the pressure gauge. Reference "Analog

Sealing" chart in this section for correct values.

If the transducer does not produce the proper

pressures, check for air leaks in the air pilot tube,

transducer and fluid regulator. The transducers

are very low volume devices and are designed

for non-flow, or dead-headed operation. Also,

observe the pressure gauge for oscillation and

listen to the transducer for fluttering sounds. These

are other indications that there may be air leaks

in the system.

Another method for isolating small air leaks is to

spray or dab a small amount of a soap and water

solution around the fittings or valves and look for

the continuous formation of air bubbles.

Advertising