Connecting and setting up the peristaltic pump

A peristaltic pump 772 Pump Unit is used for aspirating the rinsing station
at tower 2.

6.1812.000

6.1812.000

6.1820.050

6.1826.100

1

2

3

4

1 Cut one piece of 6.1826.100 pump tubing down to a length of

approx. 17 cm and attach a 6.1820.050 tubing olive at both ends.

2 Fasten a piece of 6.1812.000 PTFE tubing leading to a waste con-

tainer to one of the tubing olives. Attach the 6.1812.000 PTFE tubing
that is connected to the rinsing station as a drain connection to the
other tubing olive.

3 Release the locking lever by rotating the clamping screw and open

the pressure clamp.

4 Loop the pump tubing around the rotor and fasten with the tubing

clamps. The rotor turns in clockwise direction. The inlet tubing must
therefore be clamped below, the outlet tube above.