4 configuring the towers, Configuring the towers – Metrohm 815 Robotic Soliprep User Manual

Page 21

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Robotic Titration Soliprep 2.815.1110

815 Robotic Soliprep

■■■■■■■■

13

Swing direction = +

5 Finish the settings with [OK].

The instrument must be reinitialized in order to make the settings
effective.

6 Click [OK] on the tab.

7 Right-click on the instrument name 815_1 in the Devices window

and click on Initialize.

The Sample Processor is initialized. The settings of the Swing Heads
and robotic arms are now valid.

2.1.4

Configuring the towers

Various positions must be defined for lift 1 and lift 2 before process meth-
ods are created for the 815 Robotic Soliprep.

Tower 2

Configure the positions for tower 2:

1 Click on the Tower 2 tab.

2 Enter the following settings:

Swing position = 29 mm

External position 1, Angle [°] = 119.6

External position 1, Work position [mm] = 157