15 mounting the rinsing tubings for the polytron, 16 connecting the polytron, Mounting the rinsing tubings for the polytron – Metrohm 815 Robotic Soliprep User Manual

Page 118: Connecting the polytron

Advertising
background image

4.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

110

■■■■■■■■

815 Robotic Soliprep

Mount the Polytron as follows:

1 Insert the Polytron with the aggregate already mounted into the

robotic arm holder on tower 2 from above.

2 Readjust the position of the holder if necessary. Press down the fixing

lever of the holder to accomplish this.

4.1.15

Mounting the rinsing tubings for the Polytron

®

For rinsing the Polytron aggregate rinsing nozzles must be mounted on
the robotic arm. Proceed as follows:

1

2

2

2

1

1

3

6.2740.020

6.1805.060

6.1805.xxx

1 Replace the three stoppers on the Polytron robotic arm with

6.2740.020 rinsing nozzles.

2 Connect the three rinsing tubings that are connected to the tower 2

distributor to the rinsing nozzles.

3 The preinstalled tubing of the robotic arm can be used to add solvent

to the sample.

4.1.16

Connecting the Polytron

®

The Polytron

®

PT 1300 D is comprised of a control module and a dis-

persion drive with dispersion aggregate for mixing and shredding solid
samples. The control device is directed by a PC software, e.g. tiamo,
via a serial RS-232 connector.

Please refer to the operating instructions of the Polytron for details regard-
ing connecting. The Polytron is connected as follows: