17 equipping the titration head, Equipping the titration head – Metrohm 815 Robotic Soliprep User Manual
Page 39
Advertising
■■■■■■■■■■■■■■■■■■■■■■
2 Robotic Titration Soliprep 2.815.1110
815 Robotic Soliprep
■■■■■■■■
31
1
3
4
2
5
1 Screw a filled adsorber tube onto the Vent connector of a dosing
unit.
2 Provide the dosing unit with a filling tubing (Port 2 on the underside
of the dosing unit) and screw it onto a reagent bottle.
3 Place the reagent bottle with the dosing unit onto the Titrando.
4 Connect a dosing tubing (e.g. 6.1805.130) to Port 3. This tube deliv-
ers the titrant to the titration head on tower 1.
5 Place a Dosino (dosing drive) on the dosing unit. Connect the Dosino
connection cable to an MSB connector on the Titrando.
2.1.17
Equipping the titration head
Mounting the aspiration and rinsing tubing
Proceed as follows: