9 mounting the polytron robotic arm, 10 washing station and drip pan, Mounting the polytron robotic arm – Metrohm 815 Robotic Soliprep User Manual

Page 29: Washing station and drip pan

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Robotic Titration Soliprep 2.815.1110

815 Robotic Soliprep

■■■■■■■■

21

2.1.9

Mounting the Polytron robotic arm

1

6.1462.250

Mount the 6.1462.070 robotic arm to tower 2 as follows:

1 Hold the robotic arm in such a way that the holder faces to the left

and and slip it over the guide pins of the drive disc from below.
While doing so, let the robotic arm point outwards as far as possible,
i.e. towards the tower - see above.

Note

Take care to ensure that you do not twist the drive disc, thus caus-
ing pressure against the drive.

2 Screw the robotic arm to the Swing Head tightly with the screws and

washers provided.

2.1.10

Washing station and drip pan

To mount the washing station and the drip pan, remove the sample rack.
Now proceed as follows:

1 Mount the washing station to the left next to tower 2 on the assem-

bly rail and screw it tightly.

Advertising