2.1 Installation
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
815 Robotic Soliprep
2.1.5
Defining rack positions
Rack 6.2068.030
1 Switch over to the Rack tab and click on [Rack data] to open the
rack parameter settings.
2 Switch to the Lift position tab.
3 Enter the value 150 as Work position under Tower 1.
4 Enter the value 45 mm as Shift position under Tower 1.
5 Enter the value 159 mm as Work position under Tower 2.
6 Enter the value 29 mm as Shift position under Tower 2.