4 setting lift positions, Setting lift positions – Metrohm 815 Robotic Soliprep User Manual

Page 128

Advertising
background image

4.4 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

120

■■■■■■■■

815 Robotic Soliprep

2 Double-click on the instrument name 815_1.

3 Switch over to the Rack tab and open the rack parameter settings

with a click on [Rack data].

4 Correct the value for Rack offset according to the observed devia-

tion from the positioning reticle. One tick mark corresponds to
approx. 0.5°.

5 Close both the rack data configuration and the properties dialog of

the Sample Processor with [OK].

6 In the manual control, select the same rack position again and lower

the lift down to the positioning reticle.

Now the adapter tip should point to the middle of the positioning
reticle. If this is not the case, then an additional correction must be
made.

4.4

Setting lift positions

Some lift positions must be adjusted precisely in order to ensure a perfect
automation sequence. Use the manual control in tiamo for this pur-
pose. The following section describes how you can approach the individ-
ual positions and adjust them as needed.

The sample rack must be charged with all of the inserts.

Lift positions as method variables
The required lift positions that have not been determined in the 815
Robotic Soliprep configuration (see Chapter 4.1.4, page 91) are defined as
method variables in the method 815 Robotic Filtration Soliprep -
standard method
.