3 Robotic Flexible Soliprep 2.815.2110

815 Robotic Soliprep

■■■■■■■■

53

2 Inserting into the guide chain

Place the required tubings and cables into the guide chain.

3 Closing the guide chain

Close the clip of each chain link again by hand and apply forceful
pressure to snap them into place.

Caution

Take care to ensure that any cables and tubings that are fixed in the
guide chain are not in a position to hinder the free movement of the lift
or of the robotic arm. This could lead to damage to the affected drive.

Remove the clips of the three lowest chain links by loosening them on
both sides.

3.1.8

Mounting the deflector and the collection container

When injection needles and disposable filters are used, these need to be
stripped off the robotic arm again afterwards. A deflector has been moun-
ted for this purpose. Proceed as follows:

Mounting the deflector

1 Loosen the uppermost two screws on the right-hand side of tower 1.

2 Screw the 6.2058.050 fastening plate tightly with the aid of the

screws supplied.