11 connecting and setting up peristaltic pumps, Connecting and setting up peristaltic pumps – Metrohm 815 Robotic Soliprep User Manual

Page 32

Advertising
background image

2.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

815 Robotic Soliprep

2.1.11

Connecting and setting up peristaltic pumps

The two 772 Pump Unit peristaltic pumps are used for aspirating the titra-
tion vessels and the rinsing station.

772 Pump Unit for the rinsing station

6.1812.000

6.1812.000

6.1820.050

6.1826.100

1

2

3

4

1 Cut one piece of 6.1826.100 pump tubing down to a length of

approx. 17 cm and attach a 6.1820.050 tubing olive at both ends.

2 Fasten a piece of 6.1812.000 PTFE tubing leading to a waste con-

tainer to one of the tubing olives. Attach the 6.1812.000 PTFE tubing
that is connected to the rinsing station as a drain connection to the
other tubing olive.

3 Release the locking lever by rotating the clamping screw and open

the pressure clamp.

4 Loop the pump tubing around the rotor and fasten with the tubing

clamps. The rotor turns in clockwise direction. The inlet tubing must
therefore be attached below, the outlet tube above.