15 connecting the polytron, Connecting the polytron – Metrohm 815 Robotic Soliprep User Manual

Page 38

Advertising
background image

2.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

815 Robotic Soliprep

2 Connect the three rinsing tubings that are connected to the tower 2

distributor to the rinsing nozzles.

3 The preinstalled tubing of the robotic arm can be used to add solvent

to the sample.

2.1.15

Connecting the Polytron

®

The Polytron

®

PT 1300 D is comprised of a control module and a dis-

persion drive with dispersion aggregate for mixing and shredding solid
samples. The control device is directed by a PC software, e.g. tiamo,
via a serial RS-232 connector.

Please refer to the operating instructions of the Polytron for details regard-
ing connecting. The Polytron is connected as follows:

1

Connect the connecting cable of the dispersion device to the
front of the control module.

2

Check the set supply voltage on the connection socket on the rear
side of the control module.

Switch off the instrument.

Plug in the power supply cable and connect to a socket.

3

Connect the 6.2134.110 RS-232 connection cable to the 9-pin
connection socket on the rear side of the control module.

Plug in the other end of the cable to the COM1 or COM2 port on
the PC. The port on the PC is usually marked with IOIOI.
If the PC has no serial interface, then a USB port on the PC can be
used with the aid of an RS-232/USB converter (e.g. the
2.145.0320 Edgeport).

Note

The Polytron must be registered manually in tiamo as an RS-232
device.

2.1.16

Installing a titrator with titrant and auxiliary solutions

A Titrando or a Titrino model can be used as titrator for the determination
of samples. Directions for the installation of a titrant and/or an auxiliary
solution on an 836 Titrando are given below as an example:

Advertising