13 setting up the rinsing tubing, 14 assembling and mounting the polytron® drive, Setting up the rinsing tubing – Metrohm 815 Robotic Soliprep User Manual

Page 117: Assembling and mounting the polytron, Drive

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Robotic Filtration Soliprep 2.815.3110

815 Robotic Soliprep

■■■■■■■■

109

4.1.13

Setting up the rinsing tubing

Rinsing tubing at tower 2

6.1805.060

6.1805.120

1

2

Install the tubing for rinsing the rinsing station as follows:

1 Connect the green inlet tubing (6.1805.120) of the rinsing station to

one of the four connectors on the tower 2 distributor.

2 Connect three pieces of 6.1805.060 tubing (60 cm) to the tower 2

distributor.

3 Connect a 6.1812.000 PTFE tubing to the tower 2 valve. Connect the

other end of the tubing to a canister with rinsing liquid (e.g. water).

4.1.14

Assembling and mounting the Polytron

®

drive

Please consult the Polytron user manual for details how to mount the
aggregate to the Polytron PT 1300 D.

Mounting the Polytron

1