13 setting up the rinsing tubing, 14 assembling and mounting the polytron® drive, Setting up the rinsing tubing – Metrohm 815 Robotic Soliprep User Manual
Page 117: Assembling and mounting the polytron, Drive
■■■■■■■■■■■■■■■■■■■■■■
4 Robotic Filtration Soliprep 2.815.3110
815 Robotic Soliprep
■■■■■■■■
109
4.1.13
Setting up the rinsing tubing
Rinsing tubing at tower 2
6.1805.060
6.1805.120
1
2
Install the tubing for rinsing the rinsing station as follows:
1 Connect the green inlet tubing (6.1805.120) of the rinsing station to
one of the four connectors on the tower 2 distributor.
2 Connect three pieces of 6.1805.060 tubing (60 cm) to the tower 2
distributor.
3 Connect a 6.1812.000 PTFE tubing to the tower 2 valve. Connect the
other end of the tubing to a canister with rinsing liquid (e.g. water).
4.1.14
Assembling and mounting the Polytron
®
drive
Please consult the Polytron user manual for details how to mount the
aggregate to the Polytron PT 1300 D.
Mounting the Polytron
1
2