Metrohm NIRS XDS Process Analyzer – MicroBundle User Manual

Page 46

Advertising
background image

44

▪▪▪▪▪▪▪

Slide the knurled collar up onto the connector
housing. The final assembly is shown.

To attach to the instrument, insert the connector
jack (tab down) into the plug on the side of the
instrument.

Push until it “clicks” into place.

Thread the knurled collar onto the threaded
portion of the mating assembly. Tighten hand-
tight.

Be sure the knurled collar is tightened enough to
compress the O-ring, to seal out contaminants.

Purged Unit Instrument RJ-45 Connection

The Purged Enclosure uses different AC Power
and communication entrances.

The instrument air is routed to the entrance on
the filter, as shown.

AC (“mains”) power is provided to the lower
entrance on the box where shown in this photo.
This shows the temporary AC power cord. A
permanent installation must use dedicated
wiring, in accordance with local codes.

The RJ-45 fiber optic communication lines enter
the enclosure. From here, the cables go to the
fiber optic modem located inside the instrument.