2 lamp replacement, Lamp replacement – Metrohm NIRS XDS Process Analyzer – MicroBundle User Manual

Page 98

Advertising
background image

96

▪▪▪▪▪▪▪

11.2

Lamp Replacement

Lamp replacement is quite simple with the XDS Process Analytics Instrument. Lamp replacement
should normally be scheduled at factory shutdown intervals, to avoid unexpected downtime during
peak production. Lamp instructions are as follows:

1.

Shut off AC Power to the instrument. If

the location has a “Lockout Policy” it
must be followed.

2.

Using a large, flat-bladed screwdriver,

open the door latches. Turn counter-
clockwise to open. When open, the slot is
vertically-oriented.

A coin may be used if a large screwdriver
is not available.