3 instrument calibration, transmission probes, Instrument calibration, transmission probes – Metrohm NIRS XDS Process Analyzer – MicroBundle User Manual

Page 87

Advertising
background image

▪▪▪▪▪▪▪

85

6.

Vision requests that the correction file be

located. It is in the C:\Vision directory.

Highlight the file “TSSZERO.da” and click
“Open.”

7.

Once the file is selected, Vision applies the correction to sample spectra taken with this

channel. This method enhances transferability of samples from one channel (of a given probe
geometry) to another.

Click “Close Report” to proceed.

8.

Select the next channel, if used. From the

menu bar, click on Acquire, then Select
Channel. Use the spinner to select the
correct channel, then click on “OK”.

This must be performed for each channel.
Always verify the channel before
proceeding with the test.

10.4.3

Instrument Calibration, Transmission Probes

First select the channel. From the menu bar, click on
Acquire, then Select Channel. Use the spinner to select the
correct channel, then click on “OK”.

This must be performed for each channel. Always verify
the channel before proceeding with the test.