2.

Tighten the thumbscrew gently on the

probe.

3.

With both probes in position, cover the

center area shown in the photo to prevent
stray light from affecting the Reference
Standardization.

Do not insert the WST3Wxxx standard.
Nothing is to be placed between the fiber
bundle ends during Reference
Standardization.

4.

Select the channel. From the menu bar,

click on Acquire, then Select Channel. Use
the spinner to select the correct channel,
then click on “OK”.

This must be performed for each channel.
Always verify the channel before
proceeding with the test.

5.

Select Diagnostics,

Reference Standard,
Create Reference
Standard.

Vision takes an
instrument reference
scan.

The XDS instrument
has an internal
reference fiber, which
is selected for this
operation.