Configuration

39

CFGPWD

0
0–9999999

reset=999999

Configuration password. Specify a non-zero value to restrict access to all configuration
menus and PCE.

Note

If a configuration password is specified, be sure to record the password
and keep it in a safe place. If the configuration password is lost or
unavailable, the indicator must be cleared (reset configuration and
calibration parameters) to continue use.

To clear the indicator, press the setup switch, then enter 999999 (six nines) at the
configuration password prompt. The unit performs a RESETCONFIGURATION then
returns to the setup mode configuration display.

KYBDLK

OFF
ON

Keyboard lock. Specify ON to disable the keypad in normal mode.

ZERONLY

OFF
ON

Zero key only. Specify ON to disable all front panel keys except ZERO in normal mode.

SOFTKEYS

None
<blank>
Time/Date
Display Tare
Display Accum
Display ROC
Alibi

Softkeys provide additional operator functions for specific applications. Assignments are
listed on the tabs shown at the bottom of the LCD display; functions are activated by
pressing the arrow keys below the softkey tabs.

Setpoint
Batch Start
Batch Stop
Batch Pause
Batch Reset

These features are only available with the 720i Batching version software.

Weigh
Reports
Management

These features are available with the Advanced Truck Mode Only.

Weigh In
Weigh Out
Truck Regs

These features are available with the Truck Modes 1-6 Only.

SKUD

1-10

User-programmable softkeys; defined by

PCEE

.

REGULAT

NTEP
CANADA
INDUST
NONE
OIML

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater

than zero. NONE allows tares to be acquired at any weight value.

• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight

is at no load. NONE allows tares to be cleared at any weight value.

• NTEP and OIML modes allow a new tare to be acquired even if a tare is already

present. In CANADA mode, the previous tare must be cleared before a new tare can
be acquired.

• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net

mode as long as the current weight is within the specified ZRANGE. In OIML mode,
the scale must be in gross mode before it can be zeroed; pressing the ZERO key in
net mode clears the tare.

• INDUST provides a set of subparameters to allow customization of tare, clear, and

print functions in non-legal-for-trade scale installations. See Level 4 parameter
descriptions in Table 3-8 on page 41.

The value specified for this parameter affects the function of the front panel TARE and
ZERO keys. See Section 11.2 on page 101 for a complete description of TARE and ZERO
key functions for each of the regulatory modes.

REGWORD

GROSS
BRUTTO

Sets the term displayed when weighing in gross mode. Selecting BRUTTO replaces the
Gross annunciator with Brutto.

FEATURE Menu

Parameter

Choices

Description

Table 3-7. FEATURE Menu Parameters (Continued)