Calibration

49

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used
t o h a n g t e s t w e i g h t s , t h e n s e n d t h e
SC.REZERO#1 serial command.

7. Send the KSAVE serial command to save the

calibration changes; send KEXIT to exit setup
mode.

4.4

Revolution Calibration

The

Revolution III

Calibration Wizard provides

step-by-step scale calibration. With the

720i

connected

to the PC, select the scale and click on the scale icon in
the Calibration Wizard from the Tools menu on the

Revolution III

Scales display, then follow the steps

listed below to calibrate the scale. See Section 5.0 on
page 50
for a general overview of the

Revolution III

utility.

1. On the first Calibration Wizard display (see

Figure 4-3), select whether you are performing
a standard (zero and span) calibration or a
multi-point linear calibration. Click the

Next

button to continue.

Figure 4-3.

Revolution III Calibration Wizard Display

2. Enter the test weight used to calibrate the scale.

If chains or hooks are used to hold the weights,
check the box below the test weight value
entry. This adds a rezero step to the calibration
sequence.

3. Remove all weights from the scale. If chains or

hooks are used, place them on the scale. Press
the

Calibrate Zero

button to perform the zero

calibration. A message box appears when the
process is complete.

4. Apply the test weights to the scale. Press the

Calibrate Span

button to perform the span

calibration. A message box appears when the
process is complete.

5. If the option for chains or hooks was selected

in step 2, the Rezero display is shown. Remove
all weights from the scale, including chains or
hooks. Press the

Re-Zero

button to calibrate the

zero offset.

6. If performing a multi-point linear calibration,

up to five other calibration weight values can
be entered on the display. The weights must be
in ascending order and must not include zero
or the span weight. Enter the weight values and
click the

Go

button to calibrate each point.

7. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.