Calibration

47

4.0

Calibration

The

720i

can be calibrated using the front panel, serial commands, or

Revolution III

®

. Each method consists of the

following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights using hooks or chains

The following sections describe the calibration procedure for each of the calibration methods.

SCALES

SERIAL

FEATURE

PFORMT

DIG I/O

ALGOUT

VERS

WZERO

WVAL

WSPAN

WLIN

REZERO

POINT 4

POINT 5

POINT 1

POINT 2

POINT 3

GRADS

CALIBR

SCALE 1

Figure 4-1. Calibration (CALIBR) Submenu

4.1

Gravity Compensation

Gravity compensation for latitude and elevation is available for the

720i

. To calibrate with gravity compensation,

the LOCALE parameter under the FEATURE menu must be set ON, and the LATUDE (latitude) and ELEVAT
(elevation, in meters) parameters set before calibrating the indicator (see Figure 3-10 on page 32).

If the indicator is later installed at a different location, gravity compensation can be applied to a pre-calibrated
indicator by adjusting the LATUDE and ELEVAT parameters.

4.2

Front Panel Calibration

The CALIBR submenu (under the SCALES menu, see
Figure 4-2) is used to calibrate the

720i

.


WZERO

+ / -

1440

Last Zero

Calibrate

MV

Figure 4-2. WZERO Calibration Display

The zero, span, and linear calibration point displays
provide a set of softkeys used specifically for
calibration procedures:

+/–

Toggles to allow entry of negative or
positive values

Last Zero

Recalls the last established zero value to
allow calibration without removing tests
weights or product from scale.

Calibrate

Performs calibration for the selected
point

Millivolts

(or

Counts

)

Toggles between display of captured A/
D counts and captured millivolts values;
allows entry of calibration values in mV
or counts

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

Scale Configuration

) and remove all weight

from the scale platform. If your test weights
require hooks or chains, place the hooks or
chains on the scale for zero calibration.

2. With the

SCALES

menu highlighted, press the

Down

key, then select the scale to be calibrated.

P r e s s

D o w n

a g a i n (

G R A D S

p a r a m e t e r

highlighted), then press

Left

to

highlight the

CALIBR submenu (see Figure 4-1). Press Down to
go to zero calibration (WZERO). See Figure 4-2 on
page 47

.