Rice Lake 720i Programmable Indicator/Controller - Installation Manual User Manual

Page 46

Advertising
background image

40

720i Installation Manual

HEARTBEAT

OFF, 0.5—4.0

Specifies, in seconds, the rate of the heartbeat signal sent from the host to the display
module. The heartbeat is used to verify active communications between the host and the
display and to ensure that the displayed data is current.

If specified, the display module input heartbeat (HB1:I parameter) must also be set,
typically to a value slightly longer than the host HEARTBEAT parameter. (For example,
HEARTBEAT=1.5, HB1:I=2.0.) See Section 11.11 on page 110 for more information about
configuring the display module.

LOCALE

OFF
ON

Locale. Set this parameter ON to enable the LATUDE and ELEVAT parameters. Specifying
latitude and elevation of the scale site provides compensation for gravitational effects.
Attached scales must be recalibrated after changing this parameter from OFF to ON.

LATUDE

45
0–90

Latitude. Specify the latitude of the scale site in degrees. This parameter is shown only if
LOCALE=ON.

ELEVAT

345
±0–9999

Elevation. Specify the elevation of the scale site in meters. Valid values are –9999 through
9999. This parameter is shown only if LOCALE=ON.

Level 3 submenus

DATEFMT

MMDDYYYY
DDMMYYYY
YYYYMMDD
YYYYDDMM

Specifies the format used to display or print the date.

DATESEP

SLASH
DASH
SEMI

Specifies the date separator character.

TIMEFMT

12HOUR
24HOUR

Specifies the format used to display or print the time.

TIMESEP

COLON
COMMA

Specifies the time separator character.

SHIFT 1 - 3

SHIFT # START
SHIFT # END

Specifies the shift start and end time in 24 hour format (HHMMSS)

DAY START
DAY END

HHMMSS
HHMMSS

Specifies the day start and end time in 24 hour format (HHMMSS)

FEATURE Menu

Parameter

Choices

Description

Table 3-7. FEATURE Menu Parameters (Continued)