3.2 Guidelines for safe lifting and transportation

After operation, please observe the guidelines for temporarily decommissioning the unit (chap. 15.2).

CAUTION

Sliding or tilting of the unit.
Damage to the unit.

Do NOT lift or transport the unit using the door or the handle.

Transport the unit in its original packaging only.

For moving or shipping, secure the oven with transport straps.

Lift the unit at the four lower corners with the aid of 4 people and place it on a transport

pallet with wheels. Push the pallet to the desired site and then lift the unit at its four

lower corners from the pallet with the aid of four people.

Permissible ambient temperature range during transport: -10 °C / 14 °F to +60 °C / 140 °F.

You can order transport packing and pallets for moving or shipping purposes from BINDER service.

3.3 Storage

Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-

rary decommissioning (chap. 15.2).

Permissible ambient temperature range during storage: -10 °C / 14 °F to +60 °C / 140 °F.

Permissible ambient humidity: max. 70 % r.H., non-condensing

When after storage in a cold location you transfer the unit to its warmer installation site, condensation

may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and

is completely dry.

3.4 Location of installation and ambient conditions

Set up the safety drying oven FDL on a flat, even and non-flammable surface, free from vibration and in a

well-ventilated, dry location and align it using a spirit level. The site of installation must be capable of

supporting the unit’s weight (see technical data, chap. 17.3). The chambers are designed for setting up

inside a building (indoor use).

CAUTION

Danger of overheating.
Damage to the unit.

Do NOT set up units in non-ventilated recesses.

Ensure sufficient ventilation for dispersal of the heat.

Permissible ambient temperature range during operation: +18 °C / 64.4 °F up to +40 °C / 104 °F. At

elevated ambient temperature values, fluctuations in temperature can occur.

The ambient temperature should not be substantially higher than the indicated ambient tem-

perature of +25 °C / 77°F to which the specified technical data relate. For other ambient condi-

tions, deviations from the indicated data are possible.

Permissible ambient humidity: 70 % r.H. max., non-condensing.

Installation height: max. 2000 m / 6562 ft. above sea level.

When placing several units of the same size side by side, maintain a minimum distance of 250 mm / 9.84

in between each unit. Wall distances: rear 100 mm / 3.94 in, sides 160 mm / 6.29 in.