BINDER FDL 115 User Manual

Page 48

Advertising
background image

FDL (E2.1) 02/2015

page 48/75

Press key

Display 1 shows

0000

Display 2 shows PrgSelec: 2Prg10S

(1 program with max. 20 sections or

2 programs with max. 10 sections each?)

(actual setting: 2Prg10S)

Select between “2Prg10S” and “1Prg20S”

using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

Display 2 shows Prg.Time:

99:59

(max. section length 99:59 or 999:59?)

(actual setting: 99:59)

Select between 99:59 in hh:mm or 999:59 in

hhh:mm using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

Display 2 shows

Setp.sim

Ramp

(ramp or step?)

(actual setting: ramp)

Select between Ramp and Step using arrow

keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

Display 2 shows Tol.band

0

(Tolerance limits in °C)

(actual setting: 0)

Set value in °C using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

Display 2 shows Prog.Clk

Inactive

(Week program timer active or inactive?

(actual setting: Inactive)

Select between “Active” and “Inactive” using

arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

Display 2 shows 12h/24h

24h

(Display mode 12 hours or 24 hours?

(actual setting: 24h)

Select between 12 hours and 24 hours us-

ing arrow keys

Setting is shown in display 2.

Advertising