BINDER FDL 115 User Manual

Page 42

Advertising
background image

FDL (E2.1) 02/2015

page 42/75


Then define the settings for the program course. Two parameters can be set:

Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1

minute, and its maximum value is 99.59 (99 hs 59 min). If the value is 00.00, the program will start

immediately.

Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be

entered. If the program is not going to be repeated, enter the value “0”. For infinite repeats enter the

value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections.

In the last step start the selected program. These steps must be carried out subsequently.

Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before

starting a program.

Step 1 – Program selection (only with program type “2 programs” set):

Normal Display

Press program key

Display 1 shows

e.g. 1

(actual selection of the program)

Display 2 shows

SEL.PRG.

(select program 1 or 2)

Enter program number 1 or 2 using arrow

keys

Value is shown in display 1.

Next step

entry of program course settings

Press program key

Display 1 shows

e.g. 00.00

(entered delay time hh.mm)

Display 2 shows

RUN TIME

(enter delay time of program start)

Set delay time in hh.mm using arrow keys

Value is shown in display 1.

Press program key

Display 1 shows

e.g. -1

(actual selection of the number of program cycles)

Display 2 shows

REPEAT

(enter number of program cycles)

Select number of cycles -1, 0, 1 etc. using

arrow keys

Value is shown in display 1.

Last step – program start:

Press program key

Display 1 shows

e.g. 1

(selected program)

Display 2 shows

RUN PRG.

(Question: start selected program?)

Press program key

Display 1 shows

e.g. 25.5

C

actual temperature value

Display 2 shows

P01:S01 00:29:39

(time running backwards)

(actual program P01, actual section S01, and remaining

time of program section S01)

Program is running. The green LED (3d) lights up.

In addition to the green LED (3d) indicating a running program, the LED (3a) is lit if the heating is active,

or no LED if the actual temperature equals the set-point.

Advertising