11. Behavior in case of failures

11.1 Behavior after a power failure

Power failure during fixed-value operation (Normal Display): the entered parameters remain saved.

After the power returns, indicator light “AIR” (3) is lit, indicating that the heating has not yet been released

by the air flow monitoring. As an additional indication there is an acoustical signal which can be reset on

the controller. Press pushbutton “START” (4) to start prepurge. After completing prepurge and release of

the heating, operation continues with the set parameters. Now you can reset the visual indication

“RESET ALARM” on the controller.
Power failure during program operation: After the power returns, indicator light “AIR” (3) is lit, indicat-

ing that the heating has not yet been released by the air flow monitoring. As an additional indication there

is an acoustical signal which can be reset on the controller. Program course continues with the set-points

that have been reached previously during program operation. Press pushbutton “START” (4) to start pre-

purge. After completing prepurge and release of the heating, the set-points are equilibrated again. Now

you can reset the visual indication “RESET ALARM” on the controller.

11.2 Alarm messages

Alarm messages, e g. “RANGE ERROR CH1” in case of sensor rupture, are shown in Display 2 only in

Normal Display.
The message „RESET ALARM“ appears when the heating has not yet been released by the air flow mon-

itoring.
A buzzer can be activated / deactivated in the user level (chap. 10). It can be reset by pressing the EXIT

button. The alarm text shown in Normal Display goes off only if the cause of the alarm does not exist any

longer.

12. Safety devices

12.1 Temperature safety device class 2 (DIN 12880:2007)

The temperature safety device class 2 protects the unit, its environment and the charging material from

exceeding the maximum permissible temperature.
Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R

120 or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Ger-

many).
In the event of a fault in the temperature controller, the safety device (2)

permanently turns off the unit.

Only the fan continues running for safety reasons. This status is reported visually by the indicator lamp

(2a) and acoustically by the buzzer.
The operation of the safety device (2) is checked by moving it slowly counter-clockwise until it is turned

off. The safety device cut-off is reported visually by the indicator lamp (2a) and acoustically by the buzzer.
Then release the safety device by pressing the reset button (2b) and turn on the unit is as described pre-

viously.