4) At a drying temperature of 180 degrees Celsius and an average molecular weight of the solvent of

M=100g/mole. For a different drying temperature, the highest permissible solvent quantity must be re-

calculated. The calculation should be made on the basis of the "principles for the calculation of ventila-

tion of chamber dryers and continuous dryers" in accordance with industrial standard EN 1539:2009,

Appendix B.

5) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77°F and a power supply voltage fluctuation of ± 5%. The temperature data is determined in ac-

cordance to BINDER factory standard and to standard DIN 12880:2007, observing the recommended

wall clearances of 10 % of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the cabinet is fully loaded, the specified heating up times may vary according to the load.

17.4 Equipment and Options (extract)

To operate the safety drying oven, use only original BINDER accessories or accessories /

components from third-party suppliers authorized by BINDER. The user is responsible for any

risk arising from using unauthorized accessories.


Regular equipment
Electronically controlled APT.line™ heating technology
In compliance with all safety requirements according to EN 1539: 2009, EN ISO 13849-1:2008, and GUV-

R 500 Kap. 2.28 „Betreiben von Trocknern für Beschichtungsstoffe““ ("Dryers for coating materials") (for

Germany)
Multifunction program controller RD3 with digital display
Adjustable ramp functions via program editor
Temperature safety controller class 2 acc. to DIN 12880:2007
Communication and printer interface RS 422
Fresh-air monitoring with visual and audible alarms and automatic heating shut-off
Replaceable fresh-air filter cartridge (fine-particle filter for particle sizes 1 µm to 10 µm, class F6/EU6 acc.

to EN 779:2002)
FKM door gasket (for max. temp. 200 °C / 392°F)
2 chrome-plated racks
Rear exhaust duct, diameter 100 mm / 3.94 inch

Options / accessories
Access ports with various diameters, with silicone plug
Rack, chrome-plated or stainless steel
Perforated rack, stainless steel
Reinforced rack stainless steel, with 1 set rack lockings
Reinforced inner chamber with 2 reinforced racks
Lockable door
Silicone door gasket resistant to high temperatures > 200 °C / 392°F. Attention: Above 250 °C / 482°F

the gasket will age faster
Fresh air replacement filter (class F6/EU6 acc. to EN 779:2002) for particles 1µm to 10µm), with alumi-