14.5 Decontamination

The operator must ensure that proper decontamination is performed in case a contamination of the

chamber by hazardous substances has occurred.
Disconnect the chamber from the power supply prior to chemical decontamination. Pull the power plug.
Do not use decontamination agents that may cause a hazard due to reaction with components of the

device or the charging material. If there is doubt regarding the suitability of cleaning products, please

contact BINDER service.
You can use the following disinfectants:
Inner chamber

Standard commercial surface disinfectants free from acid or halides.
Alcohol based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.

For chemical disinfection, we recommend using the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from

liability by BINDER GmbH.

With every decontamination method, always use adequate personal safety controls.

In case of contamination of the interior by biologically or chemically hazardous goods, there are three

possible procedures depending on the type of contamination and charging material.
(1) The safety drying ovens FDL can be hot air sterilized at 190 °C / 374°F for at least 30 minutes. All

inflammable goods must be removed from the interior before.

(2) Spray the inner chamber with an appropriate disinfectant.

Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during

the decontamination process.

(3) If necessary, remove strongly contaminated inner chamber parts for cleaning. Sterilize them in a

sterilizer or autoclave.

In case of eye contact, the disinfectant spray may cause eye damage due to chemical

burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-

fectant spray.

Recommended precautions: To protect the eyes use sealed protective goggles.

CAUTION

Eye contact.
Eye damage due to chemical burns.

Do NOT empty into drains.

Wear protective goggles.

After using the disinfectant spray, allow the unit to dry thoroughly, and aerate it sufficiently.