2 set-point entry for program operation – BINDER FDL 115 User Manual

Page 36

Advertising
background image

FDL (E2.1) 02/2015

page 36/75


Program table corresponding to the diagram (with setting “Step”):

Section

SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

S01

40

00:30

S02

60

01:30

S03

80

01:00

S04

110

03:20

The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).

8.1.3 General notes on programming temperature transitions

If the tolerance limits set in the user level (chap. 10) are exceeded, the program is halted until the actual

temperature value returns to within the tolerance range. During this program interruption, the LED (3d)

flashes. Therefore, the duration of the program might be extended due to the programming of tolerances
The programming is saved even in case of a power failure or after turning off the unit.

After program rundown the controller returns to fixed value operation showing Normal Display and equili-

brates to the temperature value previously entered in fixed value entry mode.

Before starting the program, check the set-point value entered in fixed value entry mode.

After program rundown temperature will equilibrate to this value.

Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before

starting a program.

8.2 Set-point entry for program operation

From Normal Display, press down button X/W for 5 sec to access the program editor. Then enter the set-

points one after the other in all program sections of a selected program.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in

the user level, chap. 10).
In order to avoid incorrect programming, we recommend entering the values of the program course into a

table (template in chap. 8.3).

Example of program table (with default setting “Ramp”):

Section

SEC

Temperature set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

S01

40

00:30

S02

60

01:30

S03

90

01:00

S04

90

03:20

S05

110

00:01

The values of the program table can now be entered to the RD3 program controller.