BINDER FDL 115 User Manual

Page 47

Advertising
background image

FDL (E2.1) 02/2015

page 47/75

Press key

Display 1 shows

0000

(no function)

Display 2 shows Buzzer : Active

(setting of the alarm buzzer)

(actual setting: „Active“)

Select between “Active” and “Inactiv” using

arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Language : English

(selection of controller language)

(actual setting: English)

Select between languages German, English,

and French using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

e.g. 0004

(unit operating hours)

Display 2 shows Oper.hs 0004:28

(operating hours up to now hhhh:mm)

(actually displayed: 4 hs 28 min)

Press key

Display 1 shows

1000

(actual setting: 1000 hs)

Display 2 shows

Op.limit

1000:00

(maximum number of operating hours

hhhh:mm (actual setting: 1000 hs)

Set value using arrow keys

Setting is shown in both displays.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Op.back :

No

(reset counter of operating hours ?)

(actual setting: No)

Select between “Yes” and “No” using arrow

keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Protocol: MODBUS

(Selection of interface mode)

(actual setting: Modbus)

Select between protocols “MODBUS” and

“Printer” using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

e.g. 3

(actual setting: 3 min)

Display 2 shows Prt-Inv.

3

(print interval)

(actual setting: 3 min)

Select value between 0 and 255 minutes

using arrow keys

Setting is shown in displays 1 and 2.

Press key

Display 1 shows

0000

Display 2 shows Disp.LED:

No

(continuous display illumination?)

(actual setting: No)

Select between “Yes” and “No” using arrow

keys

Setting is shown in display 2.