Technical description, 1 factory calibration and adjustment, 2 definition of usable volume – BINDER FDL 115 User Manual

Page 65: 3 technical data fdl 115

Advertising
background image

FDL (E2.1) 02/2015

page 65/75

17. Technical description

17.1 Factory calibration and adjustment

This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using

standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-

fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of

measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 sys-

tems. They are controlled and calibrated to a DKD-Standard at regular intervals.

17.2 Definition of usable volume

The usable volume illustrated below is calculated as follows:

b

c

c

a

b

a

a

C

B

A


A, B, C = internal dimensions (W, H, D)
a, b, c = wall separation

a = 0.1*A

b = 0.1*B

c = 0.1*C

V

USE

= (A - 2 * a) * (B - 2 * b) * (C - 2 * c)

Figure 15: Determination of the usable volume

The technical data refers to the so defined usable volume.

Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them

and thus obtain a homogenous distribution of temperature and humidity.

17.3 Technical data FDL 115

Exterior Dimensions

Width

mm / inch

835 / 32.87

Height (including feet)

mm / inch

800 / 31.50

Depth

mm / inch

685 / 26.97

Additional depth of door handle

mm / inch

50 / 1.97

Wall clearance rear

mm / inch

100 / 3.94

Wall clearance side

mm / inch

160 / 6.30

Exhaust duct, outer diameter

mm / inch

100 / 3.94

Total steam space volume 1)

l / cu.ft.

156 / 5.51

Advertising