Select mode & setup – Casella CEL CEL-440 User Manual

Page 31

Advertising
background image

Operation

Press

Display Shows

Comments

11. If required,

change the

weightings to

enable

quick

edit

mode

Use

and

to select a

field.
Use

and

to change

the entry in the field.

Save the settings by

pressing

again to

resume normal operation.
Press

to return to the

Setup screen.

12. If required

select

measure-

ment

Functions

from Setup

screen

followed
by

This screen allows the

following measurement

functions to be specified

(√ =enabled, X =disabled)

using the weightings set in

step 6 above:

L

AFmx

L

Tm5

L

AFmn

L

AF10.0

L

Aeq

L

AF50.0

L

Zpk

L

AF90.0

L

AE

L

AF95.0

L

EP,v

L

AF99.0

L

Tm3

Histogram

10

When the dose

normalisation time is set to

8 hours (step 10) L

EP,v

will

be identified as L

EP,d

(Similarly, when the TWA

V

normalisation time is set to

8 hours, the unit will be

identified as TWA.)
Any L

AF

value between 0.0

and 99.9 can be set.

13. If required,

change the

functions

to

enable

quick

edit

mode

Use

and

to select a

field.
Use

and

to change

the entry in the field.

Save the settings by

pressing

again to

resume normal operation.
Press

to return to the

Setup screen.

s-14

Time Weight

F

Peak Weight

Z

Exchange Q

3

RMS Weight

A

1/4

s-15

AFmn

Aeq

L
L

L
L

Zpk

LL

AFmx

L

1/14

s-16

AFmn

Aeq

L
L

L
L

Zpk

LL

AFmx

L

L

1/14

Note 10: Although included on the display, this feature was first enabled at V 2.0 firmware.

CEL-440/480 Operators Handbook - Page 31

Select Mode & Setup