Use this option with great caution !

The option to format the instrument memory is displayed on the

final Self Test screen.

Proceed as follows.

Operation

Press

Display Shows

Comments

1. Start from

the final

Self Test

screen.

This screen offers the

Format option.

Perform Step 2 before the

option is replaced by the

Battery Voltage line and the

calibration or menu screens.

2. When

format is

required

This screen asks for

confirmation.

Use

and

to select

Yes or No, then confirm

with

.

4.5 Use With Tape and DAT Recorders

(Including Calibration For Line Input)

The sound level meters can be connected to a DAT or tape recorder and
function as an accurately calibrated input system.

This enables the measured sound (or vibration) levels to be

recorded for further calculation and analysis.

When recording, connect the recorder to the OUT Phono

jack-socket in the bottom of the sound level meter (Figure 6). This socket
supplies an unconditioned AC signal with a maximum level of 0.5 V RMS.

When replaying, insert the recorded signal via the IN Phono

jack-socket shown in Figure 7. Suitable signals from other sources may
also be inserted for measurement or analysis in this way.

If an electrical calibration of the instrument using the microphone

substitution method is required, insert the calibration signal via a
CEL-516 Line Input Adaptor (or the earlier CEL-216) screwed in to a
CEL-495 Preamplifier in place of the Type 1 microphone (which may also

f-01

512Kb

13Jan98 1Ø:54:42

1%
2
Format

Memory
Full
Runs

f-02

Delete All Runs

WARNING

and User Setups
No

Yes

Page 50 - CEL-440/480 Operators Handbook