Measure mode setup, User setup 3, Factory setup user setup 1 user setup 2 – Casella CEL CEL-440 User Manual

Page 34: Mode functions

Advertising
background image

selected preset intervals between 0 minutes and
1 hour.

When using timed scanning, make sure the repeat time is greater than
the number of channels to be scanned multiplied by the stabilization
time, otherwise some bands will not have been scanned before the scan
re-starts for the next repeat period.

For example, the maximum of 30 bands x 10 s would require a

repeat time greater than 5 minutes.

Select and save narrow band setups as follows.

Operation

Press

Display Shows

Comments

1. Select

Octave or

Third

Octave

Measure-

ment as

detailed in

Section 3.1

The highlighted option will

be the bandwidth used.

2. Display the

Main

menu, then

select

Setup

then

and

This is the Main menu

screen with the reverse

video cursor highlighting the

Setup option.

3. Move to

the Setup

selection

menu

This menu allows any of the

available setups to be

selected.

The highlighted setting will

be used.

This setting will also be

stored and used the next

time the instrument is

switched on.

4. Select a

setup and

move to

the Setup

Screen

and

then

This screen allows the

measurement Functions and

the scanning Mode to be

selected.

Calibration
Memory

s-22

Measure Mode

Setup

2/6

User Setup

3

s-23

Factory Setup
User Setup

1

User Setup

2

3/5

s-24

Mode

Functions

1/2

Page 34 - CEL-440/480 Operators Handbook

Select Mode & Setup

Advertising