Criteria 3 ensures that even when the meter is unable to reach a

steady L

eq

, it will have measured for long enough to have as good an L

eq

as the source permits, without measuring for extremely lengthy periods.

Using a typical pink noise source, autoscans take around 20

seconds for all octaves installed, and about 60 seconds for all third
octaves installed. Scanning times can be shortened by restricting the
number of bands to be measured.

A settling time for each band plus a minimum time to obtain an

adequate reading in the band are factory set in the meter and are listed
below. The maximum measurement time per band for Criterion 3 is
always fixed at 4 times the minimum measurement time.

Table 2: Octave Band Settling and Minimum Measurement Times

(Maximum Measurement Time is Four Times Minimum)

Frequency

Settling Time

Minimum

Measurement

Time

Broadband 1

640 mS

500 mS

Broadband 2

640 mS

500 mS

32 Hz

3500 mS

2 S

64 Hz

2500 mS

2 S

125 Hz

1300 mS

1 S

250 Hz

750 mS

1 S

500 Hz

400 mS

1 S

1 kHz

250 mS

500 mS

2 kHz

250 mS

500 mS

4 kHz

250 mS

250 mS

8 kHz

250 mS

250 mS

Table 3: Third-Octave Band Settling and Minimum Measurement Times

(Maximum Measurement Time is Four Times Minimum)

Frequency

Settling Time

Minimum

Measurement

Time

Broadband 1

640 mS

500 mS

Broadband 2

640 mS

500 mS

25 Hz

3500 mS

2 S

Page 64 - CEL-440/480 Operators Handbook

Specification