Both clock and RAM are also powered by an additional back-up battery,
so that neither time information nor memory contents are lost when the
main batteries are exhausted and need changing. The back-up battery
has a usable life in excess of five years, after which it must be replaced.
(Please consult CEL Service Department for details.)

The display has a matrix of 64 x 128 pixels which is sufficient for twelve
lines of information, each with nine characters. Detailed descriptions of
the various displays are included with the instructions given for each op-
eration or control option.

The condition of the main batteries is shown by a battery ikon.

indicates full battery.

indicates low battery, and will be shown when the voltage falls be-

low 3.6 V. This ikon can be removed only by switching the instrument
OFF, replacing the batteries, and restarting the measurement sequence.

The overload indication

will appear if the pre-weighting overload de-

tector is activated. The presence of an overload will also be stored with
the relevant data. Similarly, the pause indication

will appear if

is

pressed to interrupt logging and an indication will also be stored with the
relevant data.

The display is able to give an indication of the following parameters and
conditions ('though not necessarily at the same time):

Memory use,
Time weighting in use,
Frequency weighting in use,
Measurement range in use,
Current Q value,
Current period time,
Current cumulative period identity,
Current time,
Event time activity,

RUN

PAUSE

Preamplifier

RMS & Log
Conversion

Overload

Detector

Analog Sect.

Power Sup.

Digital Sect.

Power Sup.

Analog to

Digital Conv.

EPROM

RAM

Clock

Microprocessor

Battery

Pack

Memory &

Clock

Back-up Bat.

Select

On/Off

Analog

Digital

930002

Connector

18-line

Display

Linear

ON Key

A-weighting

CEL-225/3

CEL-192

CEL-230 Microphone/

Preamplifier Unit

Microphone

(CEL-268)

C-weighting

(CEL-368)

Section

Pre-

amplifier

Section

Bat. Voltage

Analog

Output

Log.

Out.

Power Sup.

Control

OFF

200 V
Polariz-

ing

Supply

Amplifier

(+30 dB)

Calibrate

10 - 80 dB

70 - 140 dB
40 - 110 dB

Cal

Keys

Figure 2: Block diagram for both instruments

Page 32 - CEL-268/368 Operator's Handbook