Casella CEL CEL-368 User Manual

Page 86

Advertising
background image

Measurement ranges - CEL-368:

Range W'ghting Display

Primary

Primary

Pulse

Setting

Range

Linearity

Linearity

Range

Range*

Range

Pk values

IEC 651

IEC 804

IEC 804

10-80

A

10-80.1 dB

23-66 dB

23-80 dB

60 dB

C

10-80.1 dB

26-66 dB

26-80 dB

57 dB

Linear

10-80.1 dB

32-66 dB

32-80 dB

51 dB

40-110 A

40-110.1 dB 40-96 dB

40-110 dB 73 dB

C

40-110.1 dB 40-96 dB

40-110 dB 73 dB

Linear

40-110.1 dB 40-96 dB

40-110 dB 73 dB

70-140 A

70-140.1 dB 70-126 dB 70-140 dB 73 dB

C

70-140.1 dB 70-126 dB 70-140 dB 73 dB

Linear

70-140.1 dB 70-126 dB 70-140 dB 73 dB

Note*

Lower reading is quoted at +10 dB on noise floor.
Upper reading allows crest factor 5.

Dynamic range:

70 dB.

Accuracy: (under reference conditions)

CEL-268: 0.7 dB,
CEL-368: 1.0 dB.

Typical instrument frequency range (within +0.5 to -3 dB)
CEL-268 - Type 1:

70-140 range: 5 Hz to 63 kHz,
40-110 range: 5 Hz to 63 kHz,
10-80 range:

5 Hz to 28 kHz.

CEL-268/368 Handbook