9.

SPECIFICATION

CEL-268 - Type 1:

Precision Sound Level Meter according to IEC 651: 1979, Type 1,
Precision Integrating Sound Level Meter according to IEC 804:
1985, Type 1,
Precision Sound Level Meter (with "R" type random incidence
microphone) according to ANSI: S1.4-1983, Type S1.

CEL-368 - Type 2:

Sound Level Meter according to IEC 651: 1979, Type 2,
Integrating Sound Level Meter according to IEC 804: 1985,
Type 2,
Sound Level Meter according to ANSI: S1.4-1983, Type S(1).

Measurement ranges - CEL-268:

Range W'ghting Display

Primary

Primary

Pulse

Setting

Range

Linearity

Linearity

Range

Range*

Range

Pk values

IEC 651

IEC 804

IEC 804

10-80

A

10-80.1 dB

26-60 dB

26-80 dB

57 dB

C

10-80.1 dB

33-60 dB

33-80 dB

50 dB

Linear

10-80.1 dB

45-60 dB

45-80 dB

38 dB

40-110 A

40-110.1 dB 40-90 dB

40-110 dB 73 dB

C

40-110.1 dB 40-90 dB

40-110 dB 73 dB

Linear

40-110.1 dB 40-90 dB

40-110 dB 73 dB

70-140 A

70-140.1 dB 70-120 dB 70-140 dB 73 dB

C

70-140.1 dB 70-120 dB 70-140 dB 73 dB

Linear

70-140.1 dB 70-120 dB 70-140 dB 73 dB

Note*

Lower reading is quoted at +10 dB on noise floor.
Upper reading allows crest factor 10.