Set Pol.*

Warning !

DO NOT connect the 200 V polarisation supply to an
electret microphone - DAMAGE WILL OCCUR.

The 200 V polarizing voltage used with condenser microphones can be
switched OFF to allow the CEL-268 to be used with an electret micro-
phone such as the MK-250.

1.

Use the SET POL. option to display the SET POL. menu.

2.

Use the only active option to switch ON or OFF.

If ON is selected, an advice menu will be displayed.

3.

Press

to switch the polarizing voltage ON only when

the condition is met by having a capacitor (=condenser) micro-
phone connected.

A confirmation message will be displayed.

SELECT

Note*:

Set Polarizing Voltage is available ONLY on the CEL-268.