Time Weighting

Fast or Slow,

Q (Exchange Rate)

4, 5, 6,
or " - " when no average is being calculated,
(Q=3 will never appear on the display, but will
always be in use for L

eq

),

Frequency Weighting

A, C, or L (Linear),

PER (Sampling Period)
i.e. the interval at which samples are taken

1 min,
5 min,
10 min,
15 min,
30 min,
60 min,

L

N

4 measurements set at any value* in 1% steps
between L

01

and L

99,

PROF

Allows profile mode to be switched ON or OFF,

EVENT

Allows event mode to be switched ON or OFF,

TIMER

Allows logging run and stop times to be set ON
or OFF.

Note*:

Period L

N

values become available only when period length 5 min.Similarly

Cumulative L

N

values become available only when run length 5 min.

Cumulative Measurement

Cumulative mode measurement for a whole run ALWAYS operations im-

mediately

is pressed to start logging, (or as soon as the timer

reaches an active start time after

is pressed).

1.

Use the control keys and the SETUP menu to select a suitable
measurement range, period, time weighting, frequency weight-
ing, four L

N

values and also a value for Q if L

Avg

is also required

for the proposed measurement task.

RUN

PAUSE

RUN

PAUSE

CEL-268/368 Handbook