Casella CEL CEL-368 User Manual

Page 88

Advertising
background image

Correction for microphone extension cable:

No significant effect with CEL extension cables up to 100 m.
With other cables, or for greater accuracy, calibrate after
installation.

Verification reference conditions:

Free field perpendicular incidence,
Ambient temperature: 20

o

C,

Measurement Range: 70 to 140 dB,
Relative Humidity: 65%,
SPL: 114 dB,
Frequency: 1 kHz.

Effect of humidity on accuracy:

Less than 0.5 dB over the range 30 to 90% RH (provided
there is no condensation), relative to the value at 65% RH and
40

o

C.

Storage range: 20 to 95% RH non-condensing.

Effect of temperature on accuracy:

Less than 0.5 dB over the range -10 to +50

o

C.

Storage range: -20 to +60

o

C.

Effect of vibration on accuracy:

Less than 6 dB when tested according to IEC 651 clause 8.3.

Effect of magnetic fields on accuracy:

Negligible at all sound levels.

CEL-268/368 Handbook

Page - 87