Fitting to targets for average speech

See Speechmap screen choices for setup and audiometric data entry. See Using Speechmap.

1. Follow the instructions in Speechmap screen choices to enter audiometric data and select the fitting rule.

2. Click

in the "2" area to start the second test. Once the test is started, click the [Stimulus] list button to

select the stimulus type. Click the [Level] list button and select a stimulus level of 65 dB. The upper bar in
this area shows the SII for the unaided stimulus at 65 dB SPL (the Unaided Audibar). When you start a new
test, the speech envelope of the previous test may be automatically hidden. See Speechmap Setup to change
this behavior. To change the default stimulus type and level see Saving test setup.

3. While the passage is being presented, adjust the hearing instrument's frequency shaping and gain for average

sounds so that the middle curve (the LTASS) falls within the target range, shown by the

symbols,

especially between 500 - 4000 Hz. These target symbols are automatically adjusted for the stimulus level you
select.

4. Click

to signal average and store a complete passage. Repeat as necessary.

5. Click

to stop the test without recording any test data.