Positioning HI for ANSI tests

1. The hearing instrument should be coupled to the appropriate blue coupler and a battery pill or battery

installed. See Coupling the hearing instrument for ANSI tests.

2. Place the instrument and coupler in the test chamber so that the microphone opening(s) of the instrument are

on the crossbar within the test zone.

3. For directional instruments, the direction of maximum sensitivity should be oriented toward the speaker.

4. Position the test box reference microphone as close as possible to the (front-facing) microphone opening of

the hearing instrument without touching it.

5. Close and seal the test chamber.

Positioning a BTE instrument for ANSI tests

Positioning an ITE instrument for ANSI tests